Test your ASTM handling with generated fixtures

In this tutorial we write a small piece of code that reads laboratory results from ASTM messages and test it with fixtures from @cosyte/synth. We start from one generated result report and end with a message that carries a non-standard escape and the warning it should raise. Every message is synthetic, so the tests can live in a public repository and print their input to a CI log.

You need @cosyte/synth and @cosyte/astm installed (Installation). Each step is a complete program you can run on its own.

1. Write the code under test​

The code under test reads an E1394 record stream with @cosyte/astm and lists each result's test code and value. Feed it a generated result report with three results. The generator returns the record stream itself:

import { parseAstmRecords, results } from "@cosyte/astm";
import { generateAstmResult } from "@cosyte/synth/astm";

function readResults(stream: string) {
return results(parseAstmRecords(stream)).map((result) => ({
code: result.universalTestId?.localCode,
value: result.value,
}));
}

const found = readResults(generateAstmResult({ seed: 12345, resultCount: 3 }));

found.length; // => 3
found.every((result) => result.code !== undefined && result.value !== undefined); // => true

resultCount sets how many R records the report holds; without it the generator picks one to four from the seed.

2. Check what the fixture carries​

A generated message says it is test data: field 12 of the H record, the processing ID, is T, never P for production. The patient's practice-assigned and laboratory-assigned IDs are minted separately, so they differ, and the stream reads back with no warnings:

import { parseAstmRecords, patient } from "@cosyte/astm";
import { generateAstmResult } from "@cosyte/synth/astm";

const stream = generateAstmResult({ seed: 12345 });
const header = (stream.split("\r")[0] ?? "").split("|");
const message = parseAstmRecords(stream);
const ids = patient(message);

header[11]; // => "T"
ids?.practiceAssignedId !== ids?.laboratoryAssignedId; // => true
message.warnings.length; // => 0

3. Read the framed form​

An analyzer on a serial line or TCP wraps the records in E1381 frames, each with a frame number and a checksum. generateAstmResultFramed frames the same records the unframed generator gives for the same seed, so the reader should find the same results once @cosyte/astm has taken the frames off:

import { parseAstmRecords, parseFramedAstm, results } from "@cosyte/astm";
import { generateAstmResult, generateAstmResultFramed } from "@cosyte/synth/astm";

const framed = parseFramedAstm(generateAstmResultFramed({ seed: 12345 }));
const unframed = parseAstmRecords(generateAstmResult({ seed: 12345 }));

framed.frameWarnings.length; // => 0
results(framed.message).length === results(unframed).length; // => true

4. Pin the seed​

A test that regenerates its fixture from a seed needs no fixture file, and its input does not change between runs, because the same seed gives the same bytes:

import { generateAstmResult } from "@cosyte/synth/astm";

generateAstmResult({ seed: 12345 }) === generateAstmResult({ seed: 12345 }); // => true

The bytes come from @cosyte/astm's builders, so pin the versions of @cosyte/synth and @cosyte/astm in your lockfile as well as the seed.

5. Add a non-standard escape​

Some instruments write escape sequences the standard does not define. generateAstmQuirk puts a &Z& escape in an otherwise clean report, and checks before it returns that @cosyte/astm raises exactly the warning the quirk is meant to cause. Your code should still read the results, and the parser should flag the escape:

import { parseAstmRecords, results } from "@cosyte/astm";
import { generateAstmQuirk } from "@cosyte/synth/astm";

const fixture = generateAstmQuirk({ seed: 12345, quirk: "unknown-escape" });
const message = parseAstmRecords(fixture.content);

fixture.intendedWarnings; // => ["ASTM_UNKNOWN_ESCAPE_SEQUENCE"]
message.warnings.map((w) => w.code); // => ["ASTM_UNKNOWN_ESCAPE_SEQUENCE"]
results(message).length > 0; // => true

What you built​

A reader for ASTM laboratory results, tested against a record stream and its framed form regenerated from a seed, plus a message with a known deviation and the warning it raises. None of it holds patient data.