Receive results from an analyzer

In this tutorial we receive one result upload the way a laboratory information system does: the analyzer opens a transfer, sends its records in frames, and closes the transfer; we answer each step, collect the records, and read the result. We play the analyzer too, so everything runs on one machine with no instrument and no connection.

You need @cosyte/astm installed (Installation) and a way to run TypeScript, such as npx tsx file.ts. Each step is a complete program that repeats what the steps before it built, so you can run any one of them on its own.

Step 1: make the bytes the analyzer sends​

An analyzer sends each record in a frame: a start byte, a frame number, the record text, an end byte, a checksum and a line end. We make those bytes with composeAstmFrames, standing in for the analyzer: a header, a patient, one result and a terminator, all synthetic.

import { composeAstmFrames, detectFraming } from "@cosyte/astm";

const wire = composeAstmFrames([
"H|\\^&|||analyzer^1\r",
"P|1|PRAC-0001|LAB-0009\r",
"R|1|^^^687|28.6|U/L|10-40|N||F\r",
"L|1|N\r",
]);

wire[0]; // => 0x02
detectFraming(wire).framing; // => "framed"

The first byte is STX (0x02), the start of a frame. detectFraming reads that leading byte and tells us the stream is framed. Some analyzers send bare records over TCP instead, with no frames; Decode a framed byte stream shows how to tell the two apart.

Step 2: decode the frames​

decodeAstmFrames splits the bytes into frames, checks each checksum and frame number, and returns the record bytes the frames carried:

import { composeAstmFrames, decodeAstmFrames } from "@cosyte/astm";

const wire = composeAstmFrames([
"H|\\^&|||analyzer^1\r",
"P|1|PRAC-0001|LAB-0009\r",
"R|1|^^^687|28.6|U/L|10-40|N||F\r",
"L|1|N\r",
]);

const { frames, records, warnings } = decodeAstmFrames(wire);

frames.map((f) => f.frameNumber); // => [1, 2, 3, 4]
frames.every((f) => f.checksum.valid && f.trusted); // => true
records.length; // => 4
warnings; // => []

On a real connection the frames arrive one at a time, and the receiver has to answer each one before the analyzer sends the next. That is the next step.

Step 3: answer the analyzer with the protocol receiver​

The transfer has three phases. The analyzer sends ENQ to ask to send, then the frames, then EOT when it is done. The receiver answers ENQ and each good frame with ACK. ltpReduce decides those answers: we give it the current state and what arrived, and it returns the new state and the actions to take. It never touches a connection itself, so the program below writes the answers into a list where a real one would write them to the serial port or socket.

import {
composeAstmFrames,
decodeAstmFrames,
ltpInitialState,
ltpReduce,
type LtpEvent,
} from "@cosyte/astm";

const wire = composeAstmFrames([
"H|\\^&|||analyzer^1\r",
"P|1|PRAC-0001|LAB-0009\r",
"R|1|^^^687|28.6|U/L|10-40|N||F\r",
"L|1|N\r",
]);
const { frames } = decodeAstmFrames(wire);

// What arrives, in order: ENQ, each frame, EOT.
const events: LtpEvent[] = [
{ type: "enq" },
...frames.map((frame): LtpEvent => ({ type: "frame", frame })),
{ type: "eot" },
];

let state = ltpInitialState();
const answers: string[] = [];
const delivered: Uint8Array[] = [];
for (const event of events) {
const step = ltpReduce(state, event);
state = step.state;
for (const action of step.actions) {
if (action.type === "deliverRecord") delivered.push(action.record);
else answers.push(action.type);
}
}

answers; // => ["sendAck", "sendAck", "sendAck", "sendAck", "sendAck"]
delivered.length; // => 4
state.phase; // => "neutral"

Five acknowledgements: one for ENQ and one for each of the four frames. Each accepted frame also produced a deliverRecord action carrying a complete record. After EOT the receiver is back in its neutral phase, ready for the next transfer.

Step 4: see a damaged frame refused​

Lines are noisy. Here the result frame arrives with one byte changed, so its checksum no longer matches. The receiver answers NAK instead of ACK, keeps nothing from that frame, and waits for the analyzer to send it again:

import {
composeAstmFrames,
decodeAstmFrames,
ltpInitialState,
ltpReduce,
type AstmFrame,
type LtpEvent,
} from "@cosyte/astm";

const wire = composeAstmFrames([
"H|\\^&|||analyzer^1\r",
"P|1|PRAC-0001|LAB-0009\r",
"R|1|^^^687|28.6|U/L|10-40|N||F\r",
"L|1|N\r",
]);
const good = decodeAstmFrames(wire).frames;

// The same bytes with "28.6" changed to "38.6" on the way.
const damagedWire = wire.slice();
damagedWire[new TextDecoder("latin1").decode(wire).indexOf("28.6")] = 0x33;
const damaged = decodeAstmFrames(damagedWire).frames[2];

function arrived(frame: AstmFrame | undefined): LtpEvent {
if (frame === undefined) throw new Error("no such frame");
return { type: "frame", frame };
}

const events: LtpEvent[] = [
{ type: "enq" },
arrived(good[0]),
arrived(good[1]),
arrived(damaged),
arrived(good[2]), // the analyzer sends the result frame again after our NAK
arrived(good[3]),
{ type: "eot" },
];

let state = ltpInitialState();
const answers: string[] = [];
const codes: string[] = [];
for (const event of events) {
const step = ltpReduce(state, event);
state = step.state;
for (const action of step.actions) {
if (action.type !== "deliverRecord") answers.push(action.type);
}
for (const warning of step.warnings) codes.push(warning.code);
}

damaged?.trusted; // => false
answers; // => ["sendAck", "sendAck", "sendAck", "sendNak", "sendAck", "sendAck"]
codes; // => ["ASTM_LTP_FRAME_REJECTED"]

The receiver never answers ACK to a frame the decoder did not trust, and never adds that frame's text to a record. The ASTM_LTP_FRAME_REJECTED warning says a frame was refused; like every warning here, it carries no record content.

Step 5: read the results​

The delivered records are the record stream. We join them, parse them, and read each message the stream carried:

import {
composeAstmFrames,
decodeAstmFrames,
ltpInitialState,
ltpReduce,
messages,
parseAstmRecords,
type LtpEvent,
} from "@cosyte/astm";

const wire = composeAstmFrames([
"H|\\^&|||analyzer^1\r",
"P|1|PRAC-0001|LAB-0009\r",
"R|1|^^^687|28.6|U/L|10-40|N||F\r",
"L|1|N\r",
]);
const events: LtpEvent[] = [
{ type: "enq" },
...decodeAstmFrames(wire).frames.map((frame): LtpEvent => ({ type: "frame", frame })),
{ type: "eot" },
];

let state = ltpInitialState();
const delivered: Uint8Array[] = [];
for (const event of events) {
const step = ltpReduce(state, event);
state = step.state;
for (const action of step.actions) {
if (action.type === "deliverRecord") delivered.push(action.record);
}
}

// Join the delivered records into one byte stream and parse it.
const stream = new Uint8Array(delivered.reduce((n, r) => n + r.length, 0));
let offset = 0;
for (const record of delivered) {
stream.set(record, offset);
offset += record.length;
}
const msg = parseAstmRecords(stream);

msg.warnings; // => []
messages(msg).length; // => 1
for (const m of messages(msg)) {
m.patient?.practiceAssignedId; // => "PRAC-0001"
m.results[0]?.value; // => "28.6"
m.results[0]?.units; // => "U/L"
m.results[0]?.status.isActiveFinal; // => true
}

We read through messages() rather than the flat results() because one transfer can carry several messages, each with its own patient. messages() keeps every result with the patient of the message that carried it. Read a stream that carries several messages explains why that matters.

What we built​

A receiver that answers the analyzer correctly, refuses a damaged frame, and reads the results only from frames the decoder trusted. To put it on a real connection, read bytes from your serial port or socket, decode each frame as it completes, feed ltpReduce the events, and write ASTM_ACK or ASTM_NAK back for each sendAck or sendNak action. Drive the LTP receiver over your own connection shows that loop, and the timers it leaves to you.