Map local test codes to LOINC

An analyzer sends its own local test code in a result's Universal Test ID, and LOINC is mapped later. applyLivd does that mapping from an IICC LIVD ("LOINC to Vendor IVD") catalog you supply. It adds an annotation beside each record, never changes the record's code or value, and never guesses a LOINC. We bundle no LOINC, SNOMED or LIVD data, so the catalog and its license obligations are yours.

Build a catalog and annotate a message​

import { applyLivd, defineLivdCatalog, parseAstmRecords } from "@cosyte/astm";

const catalog = defineLivdCatalog([{ vendorCode: "687", loinc: "1920-8", loincLongName: "AST" }]);
const msg = parseAstmRecords("H|\\^&\rR|1|^^^687|28.6|U/L||N||F\rR|2|^^^999|5|U/L||N||F\rL|1\r");
const { annotations, warnings } = applyLivd(msg, catalog);

annotations[0]?.mapping; // => { status: "mapped", loinc: "1920-8", loincLongName: "AST", source: "livd", derived: true }
annotations[1]?.mapping.status; // => "unmapped"
warnings.map((w) => w.code); // => ["ASTM_LIVD_UNMAPPED_CODE"]

applyLivd annotates each R and O record. The lookup is keyed on the vendor code alone, matched exactly and case sensitively. A code the catalog does not hold stays as sent, with an unmapped answer and a value-free ASTM_LIVD_UNMAPPED_CODE warning.

Read every outcome​

mapping.status is a closed set, so a switch over it covers every case:

statusWhat happenedWarning
mappedyour catalog vouched for exactly one LOINC for the vendor codenone
unmappedthe vendor code was looked up and your catalog held no entryASTM_LIVD_UNMAPPED_CODE
ambiguousseveral LOINCs that the reported units did not settle; all are listed, none chosenASTM_LIVD_AMBIGUOUS_MAPPING
no-vendor-codethe first component is filled and there is no vendor code, so nothing was looked upnone
no-codethe Universal Test ID carried no code at allnone

Let the catalog answer, not the wire​

The first component of the Universal Test ID is a LOINC slot, but the guide the LIVD format comes from puts sending LOINC directly from instruments out of its scope: the analyte arrives as a vendor code. So the lookup runs whenever a vendor code is present, keyed on that code, and a filled first component never answers instead of it. We perform no LOINC validation of any kind, so Glucose and 2345-7 in that slot are treated the same way: carried as unvalidatedWireValue, never reported as a LOINC, and never used as a key.

import { applyLivd, defineLivdCatalog, parseAstmRecords } from "@cosyte/astm";

const catalog = defineLivdCatalog([{ vendorCode: "687", loinc: "1920-8" }]);
const msg = parseAstmRecords(
"H|\\^&\rR|1|Glucose^^^687|28.6|U/L||N||F\rR|2|Glucose|28.6|U/L||N||F\rL|1\r",
);
const [withCode, withoutCode] = applyLivd(msg, catalog).annotations;

withCode?.mapping.status; // => "mapped"
withCode?.reportedCode; // => "687"
withCode?.unvalidatedWireValue; // => "Glucose"
withCode?.wireValueDisagreesWithCatalog; // => true

withoutCode?.mapping.status; // => "no-vendor-code"
withoutCode?.reportedCode; // => undefined

wireValueDisagreesWithCatalog says the two differ and nothing more: both values stay, neither is marked right, and neither is rewritten. It is true only where your catalog vouched for exactly one LOINC, the first component is filled, and the two are not identical; it is false everywhere else and never absent. Which source to believe is a clinical judgment we leave to you.

Choose between LOINCs by unit​

One vendor code mapping to several LOINCs is the ordinary case: the LIVD guide's own worked examples include a serum glucose reported as a mass concentration or as a substance concentration. Give each row a representativeUnit, and the units the result reports choose between them:

import { applyLivd, defineLivdCatalog, parseAstmRecords } from "@cosyte/astm";

const glucose = defineLivdCatalog([
{ vendorCode: "GLU", loinc: "2345-7", representativeUnit: "mg/dL" },
{ vendorCode: "GLU", loinc: "14749-6", representativeUnit: "mmol/L" },
]);
const msg = parseAstmRecords(
"H|\\^&\rR|1|^^^GLU|5.5|mmol/L||N||F\rR|2|^^^GLU|99|MG/DL||N||F\rL|1\r",
);
const [byUnit, noMatch] = applyLivd(msg, glucose).annotations.map((a) => a.mapping);

byUnit?.status === "mapped" ? byUnit.loinc : undefined; // => "14749-6"
byUnit?.status === "mapped" ? byUnit.unitComparison?.comparison : undefined; // => "verbatim-case-sensitive"
byUnit?.status === "mapped" ? byUnit.unitComparison?.ucumSemantic : undefined; // => false

noMatch?.status === "ambiguous" ? noMatch.reason : undefined; // => "no-candidate-matched-units"
noMatch?.status === "ambiguous" ? [...noMatch.candidates] : []; // => ["2345-7", "14749-6"]

The comparison is verbatim and case sensitive, and it is not UCUM. Nothing is normalized, case folded, scaled or converted on either side, so MG/DL does not match mg/dL, and g/L never matches mg/dL however convertible they are. UCUM asks a program claiming full conformance to compare units by their meaning; we make no such claim, and every answer a unit chose says so in unitComparison.

When a unit cannot decide, the answer is ambiguous, with every candidate listed and no LOINC chosen: the units match no candidate, match more than one, or were not reported or not readable (absent, empty, whitespace only, or lost to a malformed record). reason says which, and candidateDetails carries each candidate row so a person can choose.

A few rules about the unit attributes:

  • A candidate whose representativeUnit is absent, empty or whitespace only is never chosen by a unit, and is still listed among the candidates.
  • A vendor code with exactly one candidate LOINC is answered whether or not the units agree, with no unitComparison, because no unit chose anything.
  • On a mapped answer, representativeUnit describes the catalog row, not the record. Where several rows give the same LOINC with the unit spelled differently, the answer takes the first row's attributes. Only unitComparison says the record's units matched; branch on that.
  • vendorSpecimenDescription and vendorResultDescription are carried as written and never matched on: the LIVD guide says that text is not meant to be parsed by software that automates the mapping.

All three attributes are optional, and a catalog with none of them answers as if they did not exist.

Say which publication the mapping came from​

Four optional values describe the publication the rows came from: who published it, its version, the LOINC version the mapping was made against, and the LOINC copyright statement.

import { applyLivd, defineLivdCatalog, parseAstmRecords } from "@cosyte/astm";

const catalog = defineLivdCatalog([{ vendorCode: "687", loinc: "1920-8", loincLongName: "AST" }], {
publisher: "Example Diagnostics",
publicationVersion: "LIVD-2026-01-A",
loincVersion: "2.78",
loincCopyright:
"This content LOINC® is copyright © 1995 Regenstrief Institute, Inc. and the LOINC Committee, and available at no cost under the license at http://loinc.org/terms-of-use",
});
const msg = parseAstmRecords("H|\\^&\rR|1|^^^687|28.6|U/L||N||F\rL|1\r");

applyLivd(msg, catalog).annotations[0]?.catalogLoincVersion; // => "2.78"
catalog.publication?.publisher; // => "Example Diagnostics"
catalog.warnings?.length; // => 0

The LOINC version rides on every annotation the catalog produces, whatever the lookup answered, so a stored mapping records what it was made against. The LOINC license requires a statement of attribution and notice that LOINC content is copyrighted, and loincCopyright keeps your statement beside the mappings it covers.

Leaving out the LOINC version is allowed: the catalog is built and answers as before, and catalog.warnings holds one value-free ASTM_LIVD_CATALOG_NO_LOINC_VERSION warning.

We validate none of these four values. Each is stored as you supplied it, and catalogLoincVersion never means a LOINC was checked against that version. Carrying an attribution statement is not making one: the license obligations stay yours.

Scope the catalog to your analyzers​

The Universal Test ID carries no manufacturer, so the catalog is keyed on the vendor code alone. Two different instruments that use the same code for different analytes would both match. Build the catalog for the analyzers you actually receive from. Conflicting rows within one catalog are caught and answered ambiguous.

Write your own catalog​

LivdCatalog is an interface, so you can implement lookup against your own terminology service instead of calling defineLivdCatalog. Two rules apply to such a catalog: a lookup that throws reaches your caller unchanged, so your own failure is never read as "not in the catalog"; and a hit whose loinc is an empty string is treated as a miss.

Reference​